Ocular ultrasonography – part 1

Samuel Mauchlen BVM&S, MRCVS and Chris Dixon BVSc, CertVOphthal, MRCVS in the first of a two-part article, discuss the indications for this assessment, as well as technique and eye anatomy.

Chris Dixon, Samuel Mauchlen

Job Title



Ocular ultrasonography – part 1

The use of ocular ultrasonography has become a standard imaging technique in veterinary ophthalmology. It is used routinely to evaluate ocular and orbital structures.

Figure 1. Examples of different types of ultrasound transducers or probes. 1a. A micro-convex transducer producing frequencies of between 6MHz and 10MHz. 1b. A linear transducer producing frequencies of between 5MHz and 13MHz. 1c. A more specialised linear “hockey stick” transducer that can produce frequencies of up to 18MHz.
Figure 1. Examples of different types of ultrasound transducers or probes. Left. A micro-convex transducer producing frequencies of between 6MHz and 10MHz. Middle. A linear transducer producing frequencies of between 5MHz and 13MHz. Right. A more specialised linear “hockey stick” transducer that can produce frequencies of up to 18MHz.

In the first of two descriptive articles focusing on ocular ultrasonography, the indications, technique, and normal appearance of the eye in canine and feline patients will be discussed.

Indications for ultrasonographic assessment of the eye and orbit include:

  • examination of the intraocular structures if obscured by ocular pathology, such as corneal oedema, hyphaema or cataracts
  • detection and evaluation of intraocular pathology, such as neoplasia
  • examination of structures within the orbit and retrobulbar space
  • acquiring dimensional measurements of ocular structures

Other imaging modalities

CT and MRI are complementary techniques that can provide excellent detail of the orbital and periorbital structures, but normally require sedation or general anaesthesia1.

Both these imaging modalities can use contrast‑enhanced techniques using iodinated or gadolinium contrast agents, respectively. This can aid detection of some pathological conditions by highlighting vascular structures and changes within the vascular supply to ocular anatomy or pathological lesions2.

Figure 2a (left). A conscious patient being held in sternal recumbency for an ocular ultrasound scan. Figure 2b (right). A close-up of a patient showing direct corneal placement of a high‑frequency linear transducer.
Figure 2a (left). A conscious patient being held in sternal recumbency for an ocular ultrasound scan. Figure 2b (right). A close-up of a patient showing direct corneal placement of a high‑frequency linear transducer.

Ultrasound biomicroscopy uses very high frequency ultrasound transducers capable of producing frequencies of 50MHz or greater. This ultrasound technology is predominantly used in specialist clinical and research settings to evaluate the anterior ocular structures, including the cornea, anterior chamber and iris3.

Optical coherence tomography is a non‑invasive scanning laser that can be used to obtain very high-resolution images for ocular biometry. The machines are extremely expensive, and the latest models are capable of retinal angiography without contrast agents. Undoubtedly, refinement of the technology for the veterinary market will continue, although its usefulness may be limited to referral clinics.

The aforementioned detailed modalities are all valuable in the assessment of ocular and orbital structures, although the expense and expertise involved in each limits its availability.

Figure 3. This schematic illustration shows the orientation of the probe in a horizontal (1) and vertical (2) plane, with the corresponding position of the probe marker. Below each image, the diagrams show the movement of the probe used in each orientation to examine the ocular structures. 1a and 2a show the fanning (angulation) movement used. 1b and 2b show the sliding motion of the probe on the ocular surface.
Figure 3. This schematic illustration shows the orientation of the probe in a horizontal (1) and vertical (2) plane, with the corresponding position of the probe marker. Below each image, the diagrams show the movement of the probe used in each orientation to examine the ocular structures. 1a and 2a show the fanning (angulation) movement used. 1b and 2b show the sliding motion of the probe on the ocular surface.

In comparison, ultrasonography offers a widely available and safe imaging modality.

Preparation and technique

Given the superficial nature of the ocular structures, adjusting the ultrasound machine settings and choosing the optimal transducer (probe) from those available is important.

Medium to high‑frequency transducers capable of producing ultrasound frequencies of between 12MHz and 20MHz are ideal for imaging intraocular structures.

As ultrasound frequency increases, the resolution of the image improves, although greater attenuation of the ultrasound beam exists, reducing tissue penetration.

Transducers capable of producing higher frequencies will provide superior resolution images of the anterior ocular structures, and lower‑frequency transducers (or using a lower frequency setting) can be useful when examining the posterior globe and retrobulbar tissue.

Linear or curved (micro-convex) transducers are commonly used during ultrasonographic assessment of ocular tissues. Linear transducers are often capable of producing

higher‑frequency ultrasound waves, although the curved surface and smaller footprint of a convex probe can often be easier to manoeuvre around the eye (Figure 1).

Figure 4. A schematic illustration of the ocular anatomy.
Figure 4. A schematic illustration of the ocular anatomy.

Depth and focal zone should be altered to optimise the image quality. Depth should be set so the area of interest occupies approximately two‑thirds to three‑quarters of the screen. The focal zone represents the area of optimal lateral resolution and should be positioned at the same level as or just below the area of interest.

Ocular ultrasonography is ideally performed with patients conscious, as sedation or general anaesthesia will often lead to retraction and rotation of the globe, which can affect image quality and orientation.

Patients can be allowed to stand or sit with the head held securely by an assistant. Sternal or lateral recumbency can also be used if necessary (Figure 2).

Prior to the examination, a topical local anaesthetic – such as one to two drops of 0.5 per cent proxymetacaine – should be applied to the corneal surface.

A water-soluble acoustic coupling gel can then be applied to the probe, and the probe placed

Figure 5. By enlarging a section of this image of a normal eye, the three-layer appearance of the normal cornea can be seen.
Figure 5. By enlarging a section of this image of a normal eye, the three-layer appearance of the normal cornea can be seen.

directly on the corneal surface.

In cases with intense blepharospasm, corneal damage or globe trauma, or where the patient is uncooperative, the eye can be imaged through the skin of the closed eyelid. However, the image quality may be compromised in comparison to a direct corneal placement of the probe.

Once positioned, the probe marker can be used for orientation. The authors would recommend medial (nasal) marker positioning for horizontal plane imaging, and dorsal positioning for vertical plane imaging. From each of these positions the probe can be fanned (angled) or slid, allowing the eye to be examined thoroughly in two orthogonal planes (Figure 3).

It is important to examine both eyes as this will allow a direct comparison between the normal and affected eye.

Figure 6. This B-mode image shows the ultrasonographic appearance of some anterior ocular structures. The highly echogenic curved appearance of the anterior and posterior lens capsules can be visualised.
Figure 6. This B-mode image shows the ultrasonographic appearance of some anterior ocular structures. The highly echogenic curved appearance of the anterior and posterior lens capsules can be visualised.

Following the examination, any gel can be gently flushed from the ocular surface using warm, sterile fluid. Given the delicacy of the ocular tissues, it may be preferential to use single‑use acoustic coupling gel to reduce the risk of contamination from a larger multi‑use bottle.

The retrobulbar structures can be visualised using the aforementioned trans-corneal technique, and it is possible to acquire images by positioning the probe over the temporal fossa.

Normal appearance and anatomy

A schematic illustration of the ocular anatomy can be seen in Figure 4.

Cornea

The normal cornea appears as a curvilinear hyperechoic structure that can be split into three parallel layers using a high‑frequency transducer (Figure 5).

The anterior convex hyperechoic layer is the epithelium, the central anechoic middle layer is the stroma and the hyperechoic posterior layer is the endothelium.

Figure 7. This B-mode image shows the ultrasonographic appearance of parts of the eye, including the optic disc and some of the retrobulbar structures. The depth scale to the right of the image shows small yellow arrowheads to denote the position of the focal zones. The focal zone should be set at the level of, or just below, the area of interest. Some ultrasound machines can enable multiple focal zones, which improves the image resolution. However, this can reduce the frame rate of the display.
Figure 7. This B-mode image shows the ultrasonographic appearance of parts of the eye, including the optic disc and some of the retrobulbar structures. The depth scale to the right of the image shows small yellow arrowheads to denote the position of the focal zones. The focal zone should be set at the level of, or just below, the area of interest. Some ultrasound machines can enable multiple focal zones, which improves the image resolution. However, this can reduce the frame rate of the display.

Flattening of the corneal surface can occur if too much pressure is applied from the transducer, and an adequate application of an acoustic coupling gel must be used.

Anterior and posterior chambers

The anterior chamber is the space between the posterior surface of the cornea and the anterior surface of the iris. The posterior chamber is the narrow triangular space posterior to the iris and anterior to the lens.

Both the anterior and posterior chamber are filled with aqueous humour and are anechoic.

Iris and ciliary body

The iris is an echogenic structure that tapers towards the pupillary margin and appears continuous with the ciliary body. The ciliary processes may be visualised with higher resolution transducers.

Lens

The anterior and posterior lens capsule are highly echogenic, and on ultrasound appear as bright convex and concave lines, respectively (Figure 6).

The normal internal fibres and nucleus of the lens appear anechoic. The curved echogenic anterior capsule of the lens will cause the ultrasound waves to refract, leading to some loss of detail at edges of the posterior lens capsule – this is an artefact known as edge shadowing.

Posterior segment

The vitreous body is an anechoic region that fills the majority of the posterior segment of the globe.

Figure 8. On direct examination, the central lens of this nine‑year‑old dog appears hazy (left); this is due to the increased density of the lens nucleus refracting a greater proportion of light. On ultrasound examination of the same eye (right), the typical appearance of nuclear sclerosis can be visualised as an echogenic, curvilinear line that is present within the lens (red arrows) running parallel to the hyperechoic lens capsule.
Figure 8. On direct examination, the central lens of this nine‑year‑old dog appears hazy (left); this is due to the increased density of the lens nucleus refracting a greater proportion of light. On ultrasound examination of the same eye (right), the typical appearance of nuclear sclerosis can be visualised as an echogenic, curvilinear line that is present within the lens (red arrows) running parallel to the hyperechoic lens capsule.

In the normal eye, the posterior tissue layers of the globe are indistinguishable and appear as a single hyperechoic curved line – representing the combined layers of the retina, choroid and sclera.

The optic disc may be visible in the posterior wall of the globe and appears as a recessed, hyperechoic area (Figure 7).

The retrobulbar space includes the extraocular muscles, retrobulbar fat, and the optic nerve and vasculature. The extraocular muscles appear as thin hypoechoic lines with the more hyperechoic retrobulbar fat forming triangular areas medial to the muscle bands. The optic nerve appears as a central hypoechoic/anechoic linear area.

In middle-aged dogs (usually older than seven years of age) the nucleus of the lens will increase in density. This is known as nuclear sclerosis. Ultrasonographically, it will appear as a thin curvilinear echogenic line within the lens, running parallel with the lens capsule (Figure 8). This can be mistaken for cataract formation.

Measurement of the ocular structures (ocular biometry) can be performed using ultrasound. Traditionally, biometry was performed using amplitude or “A-mode” scanning, which displayed a simple graphical representation of a single beam of ultrasound. In this mode, the amplitudes of the returning echoes are plotted on the Y axis and time is plotted on the X axis4.

Figure 9. In this B-mode image, biometry measurements of the globe (digital calliper line 1, from the anterior cornea to the posterior globe) and lens (calliper line 2, from the anterior to the posterior lens capsule; and line 3, measuring lens width) are shown.
Figure 9. In this B-mode image, biometry measurements of the globe (digital calliper line 1, from the anterior cornea to the posterior globe) and lens (calliper line 2, from the anterior to the posterior lens capsule; and line 3, measuring lens width) are shown.

A-mode is rarely used in clinical practice, but comparisons between A-mode and B-mode (brightness mode – the standard ultrasound mode used in clinical practice) ultrasonography have been made5. B-mode measurements are now performed in patients for ocular biometry6 (Figure 9) and the repeatability of B-mode ocular biometry measurements has been described7.

Part two will focus on the appearance of a variety of pathological conditions.